Installing the lifting column 3 - 5
Siemens AG n.a. SPR2-130.841.02 Page 5 of 10 SIREMOBIL Compact L
Medical Solutions Rev. 01 07.05 CS PS 24 System Manual
Installing the C-arm and swivel brake 3
Place the new collar over the lifting column, align it centrally and glue it tight.
Three persons are required to insert the C-arm back in the bearing.
Install the swivel brake again:
- Insert the bolt from below through the strap.
- Drive the spring pin back in with its opening in the direction of the expected pressure.
Place the Belleville springs in position. The lower spring with the cone upwards, the
upper spring with the cone downwards.
Screw the stop bush in.
At first only plug on the brake lever.
Screw tight the cover plates under the swivel brake.
Fasten the railing handle back on the C-arm.