Installing the lifting column 3 - 5
Siemens AG n.a. SPR2-130.841.02 Page 5 of 10 SIREMOBIL Compact L
Medical Solutions Rev. 01 07.05 CS PS 24 System Manual
Installing the C-arm and swivel brake 3
• Place the new collar over the lifting column, align it centrally and glue it tight.
• Three persons are required to insert the C-arm back in the bearing.
• Install the swivel brake again:
- Insert the bolt from below through the strap.
- Drive the spring pin back in with its opening in the direction of the expected pressure.
• Place the Belleville springs in position. The lower spring with the cone upwards, the
upper spring with the cone downwards.
• Screw the stop bush in.
• At first only plug on the brake lever.
• Screw tight the cover plates under the swivel brake.
• Fasten the railing handle back on the C-arm.