Installing the lifting column 3 - 9
Siemens AG n.a. SPR2-130.841.02 Page 9 of 10 SIREMOBIL Compact L
Medical Solutions Rev. 01 07.05 CS PS 24 System Manual
Adjustment of the swivel brake 3
Turn the C-arm horizontally (Fig. 16)
Measure the braking forces with a spring balance (20N) for released brake and spring
balance (200N) for applied brake at the railing handle end of the image intensifier.
In this case hook the spring balance to the railing with a cloth band or similar. Set the
brake to the above values by adjusting the stop of the brake lever.
Set the brake lever in the end positions (brake applied or released) in each case parallel
to the horizontal arm.
Apply the brake.
Remove the brake lever.
Affix the plastic guard centrally with some silicone adhesive.
Fit the brake lever in the same position as previously and screw it tight.
Remaining work 3
Glue on the new ring cover to the steering lever centrally in relation to the lever.
Switch on the system and check all functions.
Move the lifting column fully upwards and glue on the supplied scale sticker vertically in
the middle of the inner segment.
Install all covers; connect the protective conductors to the covers and reconnect the
emergency stop button.
Adjustment values:Brake released e 10 Nm Brake applied E 100 Nm