SIREMOBIL Compact L n.a. SPR2-130.841.02 Page 2 of 10 Siemens AG
System Manual Rev. 01 07.05 CS PS 24 Medical Solutions
3 - 2 Installing the lifting column
Move the lifting column upwards by applying 230V (N to X7 A/1 and phase to X7 A/2)
until the lubrication bore (1/Fig. 3) is freely accessible.
Remove the plastic cap in the bore (1/Fig.3) and oil the lifting spindle through the
opening. With the oil can, fill with oil until it can be felt through the opening with your
finger.
Use Viscogen KL 300 oil, part number 73 95 353.
Close the bore again with the plastic cap.
Install the bearing of the C-arm in the lifting column (1/Fig. 4).
Three persons are required to set the C-arm in the bearing for testing the new lifting
column (Fig. 5); secure it against twisting.
Connect the plug X7A (connection cable of the lifting column):
Fig. 3 Fig. 4
Fig. 5 Fig. 6
Neutral to X7 A/1 Phase to X7 A/2 for raising Phase to X7 A/3 for lowering
Connect the emergency stop switch for moving the lifting column.
The emergency stop switch must not be pressed.
1
1
NOTE