General 1 - 5
Siemens AG n.a. SPR2-130.841.02 Page 5 of 6 SIREMOBIL Compact L
Medical Solutions Rev. 01 07.05 CS PS 24 System Manual
.
Fig. 1 Measuring circuit for measuring the protective conductor resistance in systems that are separated from the power
supply, according to DIN VDE 0751-1:2001-10, Fig. C2.
1 = System
2 = Application part (not present)
3 = Measuring arrangement (integrated in the measuring instrument)