SITRANS F M TRANSMAG 2
Operating Instructions, 11/2009, A5E00102775-07
13
Installing
4
4.1 General Information
This chapter describes how to install the flowmeter in the compact version as well as in the
remote version.
CAUTION
The device meet the requirements of IP67 degree of protection. Proper installation is a
prerequisite for compliance with this degree of protection. Please observe the instructions
in this chapter.
The maximum permissible ambient and medium temperatures must be observed at all
times.
The measuring tube should be filled exclusively with media to which the seal and lining
materials are chemically resistant. Only conductive media are permissible.
WARNING
If the surface of the sensor or sensor tube can get hotter than 50 ºC in the application, fire
protection and/or a warning sign must be provided.
CAUTION
Make sure the measuring device is protected from direct sunlight when installing it
outdoors.