5.4 Step 3: Connecting terminal box and transmitter
SITRANS F M TRANSMAG 2
Operating Instructions, 11/2009, A5E00102775-07
29
5
6
7
5
6
7
Sensor connection
Coil cable
Sensor
29
22
30
29
22
30
23
24
23
24
55
66
55
66
Magnetic field current 1
Magnetic field current 2
PE
SmartPlug connect
Supply (-6V)
Ground (0V)
Supply (+6V)
Electrode connect
Electr. 1 (EL 1)
Electr. 2 (EL 2)
Reference coil
Ref. 1
Ref. 2
Brown
White
Black
Yellow
Green
Red
Blue
Brown
White
Green/Yellow
Electrode line
Magnetic current line
Figure 5-5 Connection diagram for power supply and signal cables on transmitter
Note
The sensors have the same terminal names, the lines must be connected 1:1.
With sensors with IP68 degree of protection, the cable colors or the identification tag on
the sensor cables must be observed.
Preparing the cables
Prepare the cable before connection as shown below.
10...12
mm
5...10
mm
75...80
mm
Connector sleeve
Braided screen
Figure 5-6 Preparation of the electrode and magnetic field current cable