Functions
7.5 Function Group Device Outputs
SITRANS F M TRANSMAG 2
52 Operating Instructions, 11/2009, A5E00102775-07
Profibus (Menu item 4.1)
This menu only appears in PROFIBUS devices. (7ME 5034-1xxxx).
This menu gives you the following setting options:
Bus address, menu item 4.1.1
The address 126 (factory setting) is reserved for newly connected devices and must be
changed to a smaller value at the latest after commissioning on the bus. A value between
30 and 125 is recommended because addresses below 30 are only assigned for bus
masters as a rule.
An address < 126 is only addressable when no cyclic communication is taking place and
all acyclic connections are closed.
With the PROFIBUS variable "factory_reset = 2712" the address can be reset to the
factory setting 126.
Note
Each address may only be assigned for one device on the connected bus!
Ident No., menu item 4.1.2
The device profile for the cyclic data traffic can be selected here.
You have the following options:
Profile-specific (=Ident No. 0x9740) means:
Restricted functions; i.e. only cyclic parameters according to profile 3.0 (flow and
totalizer net).
Manufacturer-specific (= Ident No. 0x80C4) means:
Full functions, i.e. cyclic parameters according to profile 3.0 and additional
manufacturer-specific parameters (e.g. totalizer forwards, totalizer reverse).
Note
The connected PROFIBUS master must be configured with a device database file
(GSD) according to the set ID No.
See also Device Database File (GSD) (Page 100)