Parameters
SITRANS F M TRANSMAG 2
Operating Instructions, 11/2009, A5E00102775-07
81
2.5.1 Flow Simulation of a flow value; affects all
outputs, totalizers, limit values and
display.
"F" flashes in the top right of the display
during simulation and "Simulation" is
displayed under device status.
Value Simulation value in % of full scale value 0 % -110% to +110%
Time Duration of simulation
After expiration of this time or with input of
"End" the normal measuring mode is
resumed
End
End
10 min
30 min
60 min
2.5.2 Dig. Out.1 Simulation of output signal at the digital
output 1
End
End
0.1 Hz
1 Hz
10 Hz
100 Hz
1 kHz,
10 kHz
Alarm on
Alarm off
2.5.3 Dig. Out.2
1
Simulation of the output signal at the
digital output 2
End
Alarm on
Alarm off
2.5.4 Analog Out.
1, 2
Simulation of the output signal at the
analog output
4 mA
0 mA
4mA
10mA
12mA
20mA
Failure signal
1)
Parameters apply only to devices with 20 mA/Hart communication and digital output 2 (7ME5034-0xxxx)
2)
Parameters apply only to devices with 20 mA/Hart communication and digital input (7ME5034-2xxxx)