In‐
dex
Parameter Name Data
Type
Access Annotation
20 MODE Un‐
signed8
Read /
write
Batching Mode
0: off, batching disabled
1: 1-stage batching
2: 2-stage batching
3: analog batching
21 INPUT_SELEC‐
TOR
Un‐
signed8
Read /
write
Batch input selection:
0: Mass Flow
1: Volume Flow
4: Standard Volume Flow
5: Fraction Flow Media A
6: Fraction Flow Media B
22 AMOUNT_DONE DS-101 Read only Current amount that is totalized since the start of the batch cycle. Includes
a quality code
23 AMOUNT_RE‐
MAINING
DS-101 Read only Amount that is missing to complete the current batch
24 RECIPE_ID Un‐
signed8
Read /
write
Active recipe that specifies the used parameter set.
Valid range 1…5
25 RECIPE_NAME String16 Read /
write
The recipe´s name
26 RECIPE_TAR‐
GET_AMOUNT
Float32 Read /
write
The recipe´s target amount. Follows the unit setting by RECIPE_TAR‐
GET_AMOUNT
27 REC‐
IPE_AMOUNT_UN
IT
Un‐
sign‐
ed16
Read /
write
Unit code that influences the AMOUNT_DONE, AMOUNT_REMAINING
and RECIPE_TARGET_AMOUNT values
28 - - -
-
29 RECIPE_
FIXED_COMPEN‐
SATION
Float32 Read /
write
Define the fixed pre-stop value. The value is subtracted from 'Amount' to
compensate for application delays in closing the valves.
Analog dosing
30 ANALOG_OUT‐
PUT
DS-101 Read only Calculated set point for analog valve
31 RECIPE_
CUR‐
RENT_CLOSED
Float32 Read /
write
Current level that is used in analog batching mode when the
analog valve shall be closed.
32 RECIPE_
CURRENT_PAR‐
TIAL_OPEN
Float32 Read /
write
Define current level to partly close the analog valve
Communication
2.10 Acyclic communication
FCT030 Profibus (From firmware 4.0)
62 Function Manual, 06/2018, A5E39931906-AB