Table 7-9 Digital output
Description Channels 2 to 4
Pulse 41.6 µs to 5 s pulse duration
Resolution 1 μs
Frequency 0 to 10 kHz, 50 % duty cycle, 120 % overscale provision
Resolution 0.2 Hz
Load < 750 Ω
Time constant (adjustable) 0 to 100 s
Active 0 to 24 V DC, 87 mA, short-circuit-protected
Passive 3 to 30 V DC, 100 mA, short-circuit-protected
Functions
• Pulse
• Frequency
• Alarm class / NAMUR status
• Alarm item
Table 7-10 Relay output
Description Channels 3 to 4
Type Change-over voltage-free relay contact
Load 30 V DC (30VAC peak), 100 mA
Functions
• Alarm class / NAMUR status
• Alarm item
7.6 Energy limitation parameters FST030
Table 7-11 Output parameters - wallmount enclosure
External DSL connec‐
tions
Sensor connections RTD connections
U
o
17.42V DC 17.43V DC 4.52VDC
I
o
459mA 452mA 13.8mA
P
o
2000mW 991mW 26.6mW
C
o
• Group IIC
• Groups IIB, III
• Group IIA
• 338nF
• 1969nF
• 8199nF
• 169.5nF
• 985nF
• 1000nF
• 4.28μF
• 139μF
• 994μF
Technical specications
7.6Energy limitation parameters FST030
SITRANS FS130 & FS230
Compact Operating Instructions, 12/2023, A5E38755607-AF 101