5. Tighten the F-connector using the tool until you meet a mechanical stop.
6. Remove the F-connector tool.
7. Mount and tighten the cable gland.
Repeat these steps for each sensor cable.
4.3 Connecting the sensor cables to the sensor
Note
The following example is shown in the Reect mount sensor conguration.
Connect sensor cables to the sensor as follows:
1. Fill connector end with the supplied grease prior to connecting.
2. Apply the supplied grease to the internal threads at the large end of the thread connector.
3. Observing the upstream and downstream orientation ⑥, connect upstream sensor cable ③
from the transmitter port (1A) to the sensor ⑤ and make connection snug.
Connecting
4.3Connecting the sensor cables to the sensor
SITRANS FS130 & FS230
40 Compact Operating Instructions, 12/2023, A5E38755607-AF