3. Insert wire.
4. Release wiring tool.
4.6.2 Connecting the SSL cable (industrial enclosure)
M12 connector version
The SSL cable is provided with M12 style stainless steel plugs.
The cable shield is physically and electrically terminated within the body of the plug.
Note
Never pull the cable by the plug - only by the cable itself.
1. Connect the external DSL to the transmitter using the supplied 4-wire cable with M12
connectors.
Note
Grounding
The SSL cable shield is only mechanically connected to the grounding terminal (PE), when
the M12 plug is correctly tightened.
Wiring sensor and transmitter (sensor terminal compartment)
A: Prepare the cable by stripping it at both ends.
PPPPPP
2UDQJH
<HOORZ
:KLWH
%OXH
Figure4-9 Cable end
B: Connecting sensor terminal compartment
1. Remove lock screw and remove lid.
2. Remove one of the blind plugs and 󹪝t cable gland.
3. Remove cap and ferrule from cable gland and slide onto cable.
4. Push cable through open gland; anchor cable with clamp bar.
Connecting
4.6Connecting industrial enclosure
SITRANS FS130 & FS230
58 Compact Operating Instructions, 12/2023, A5E38755607-AF