EasyManua.ls Logo

Siemens SITRANS FS130 - Construction

Siemens SITRANS FS130
118 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Table 7-17 Siemens SSL cable capacity
SSL cables
C
i
264.6pF/m
L
i
1.12µH/m
Table 7-18 Output parameters
RTD 5, RTD 6 (any combination of con‐
nections)
Sensor outputs, each
(up to 8)
U
o
4.52V DC 17.43V DC
I
o
13.8mA 452mA
P
o
26.6mW 991mW
C
o
Group IIC
Groups IIB, III
Group IIA
4.28μF
139.5μF
994μF
169.5nF
985nF
1000nF
L
o
Group IIC
Groups IIB, III
Group IIA
1.87mH
7.47mH
14.9mH
67μH
348μH
697μH
L
o
/R
o
Group IIC
Groups IIB, III
Group IIA
NA
NA
NA
18.0μH/Ω
72.2μH/Ω
144μH/Ω
Note
External capacitance and inductance
The C
o
and L
o
values already take into consideration the combined e󹪜ects of external
capacitance and inductance. No further reduction of external capacitance and inductance is
required.
7.8 Construction
Table 7-19 Designated use
Description Speci󹪝cation
Measurement of process medium
Fluid Group 1 (suitable for dangerous 󹪞uids)
Aggregate state: Paste/light slurry, liquid and gas
Technical speci󹪝cations
7.8Construction
SITRANS FS130 & FS230
104 Compact Operating Instructions, 12/2023, A5E38755607-AF