WARNING
Empty pipe detection
The installation has to fulfill the following limitations for usage of the empty pipe detection
function:
●
media conductivity ≥ 20 μS/cm
● length of cable at remote installation ≤ 50 (150 ft)
● special shield cable must be used
Note for MAG 1100 sizes DN 2 and DN 3:
● empty pipe detection is not available
● media conductivity ≥ 30 μS/cm
Note for MAG 5000/6000 CT (FW 3.03):
● empty pipe detection is not available
8.11 Dimensions and weight
MAG 1100
A
1
155 (6.10)133 (5.24)
M20
(½"NPT/
PG 13.5)
L
2
A
208 (8.19)
218 (8.58)
2 x M25
B
D
G
D
P
D
i
133 (5.24)
D L
A
1
B
Table 8-16 Dimensions in mm
Nominal
size DN
A
1)
B
1)
A
1
/A
2
2)
B
1
D D
i
D
i
(PFA) D
p
D
G
Weight
3)
mm mm mm mm mm mm mm mm mm mm kg
2 161 186 315 340 48.7 2 17.3 34 2.2
3 161 186 315 340 48.7 3 17.3 34 2.2
6 161 186 315 340 48.7 6 17.3 34 2.2
10 161 186 315 340 48.7 10 10 13.6 34 2.2
Technical data
8.11 Dimensions and weight
57