SITRANS P DS III/P410 (7MF4.33.. 7MF4.34.. 7MF4.35..)
42 A5E03434626-04, 05/2015
Level input
Lower measuring limit
Measuring cell with silicone oil filling
-100% of max. measuring range or 30 mbar a/3 kPa a/0.44 psi a depending
on the mounting flange
Measuring cell with inert liquid
-100% of max. measuring range or 30 mbar a/3 kPa a/0.44 psi a depending
on the mounting flange
Upper measuring limit
100% of maximum span
Start of scale value
between the measuring limits continuously adjustable
1)
Order the nominal measuring range with the order option Y01 for PROFIBUS PA or FOUNDATION Fieldbus.
7.3
SITRANS P410 input
Gauge pressure input
Measured variable
Gauge pressure
Span (continuously adjustable) or
measuring range, max. operating pressure
(in accordance with 97/23/EC Pressure
Equipment Directive) and max. test
pressure (in accordance with DIN 16086).
Span
1)
Maximum operating
pressure MAWP (PS)
Maximum test pressure
0.01 … 1 bar
1 ... 100 kPa
0.1514.5 psi
4 bar
400 kPa
58 psi
6 bar
0.6 MPa
87 psi
0.04 … 4 bar
4 ... 400 kPa
0.5858 psi
7 bar
0.7 MPa
102 psi
10 bar
1 MPa
145 psi
0.1616 bar
16 ... 1600 kPa
2.3232 psi
21 bar
2.1 MPa
305 psi
32 bar
3.2 MPa
464 psi
0.6363 bar
63 ... 6300 kPa
9.1914 psi
67 bar
6.7 MPa
972 psi
100 bar
10 MPa
1450 psi
1.6160 bar
0.16 ... 16 MPa
232321 psi
167 bar
16.7 MPa
2422 psi
250 bar
2.5 MPa
3626 psi
Lower measuring limit
Measuring cell with silicone oil filling
30 mbar a/3 kPa a/0.44 psi a
Upper measuring limit
100% of maximum span
Start of scale value
Between the measuring limits (fully adjustable)
1)
Order the nominal measuring range with the order option Y01 for PROFIBUS PA or FOUNDATION Fieldbus.
Differential pressure and flow rate input
Measured variable
Differential pressure and flow rate
Span (continuously adjustable) or
measuring range and max. operating
pressure (in accordance with 97/23/EC
Pressure Equipment Directive)
Span
1)
Maximum operating pressure MAWP
(PS)
2.5 … 250 mbar
0.2 ... 25 kPa
1.004100.4 inH
2
O
160 bar
16 MPa
2320 psi
6 … 600 mbar
0.6 ... 60 kPa
2.409240.9 inH
2
O

Table of Contents