Procedure
① Pressure transmitter ⑦ Drain valves
② Stabilizing valve ⑧ Condensate pots
③,
④
Dierential pressure valves ⑨ Dierential pressure transducer
⑤ Dierential pressure lines ⑩ Insulation
⑥ Shut-o valves ⑪ 3-way valve manifold
1. Open both the shut-o valves ⑥ at the pressure tapping point.
2. Open the stabilizing valve ②.
3. Wait until the steam in the dierential pressure lines ⑤ and in the equalizing vessels ⑧ has
condensed.
4. Partially open the dierential pressure valve ③ and the vent valve (sealing plug with vent
valve) on the positive side of the pressure transmitter until condensate escapes without
bubbles.
5. Close the vent valve (sealing plug with vent valve).
6. Partially open the vent valve (sealing plug with vent valve) on the negative side of the
pressure transmitter until condensate escapes without bubbles.
7. Close the dierential pressure valve ③.
8. Partially open the dierential pressure valve ④ until condensate escapes without bubbles.
9. Close the vent valve with blanking plug on the negative side ①.
10.Close the dierential pressure valve.
11.Open the dierential pressure valve ③ by half a revolution.
Commissioning
6.3 Application examples
SITRANS P320/P420
Compact Operating Instructions, 01/2021, A5E38874562-AC 53