Operating conditions according to IEC
61298-3 (2g normal)
Gauge pressure series
2)
Aluminum and stainless steel enclo‐
sure
Di󹪜erential pressure series
1)2)
Aluminum and stainless steel enclo‐
sure
Vibrations (sine)
IEC 60068-2-6
10 ... 58 Hz at 0.3 mm
58 ... 1000 Hz at 20 m/s²
1 octave/min
20 cycles/axis
1) Without mounting bracket
2) With mounting bracket
Operating conditions according to IEC
61298-3 (5g enhanced)
Di󹪜erential pressure series
1)
Aluminum and stainless steel enclosure
Vibrations (sine)
IEC 60068-2-6
10 ... 58 Hz at 0.7 mm
58 ... 1000 Hz at 50 m/s²
1 octave/min
20 cycles/axis
1) Without mounting bracket
Operating conditions for marine
(only
in combination with order option E):
Gauge pressure series
1)2)
Di󹪜erential pressure series
3)
DNV-GL (Det Norske Veritas/Germanisch‐
er Lloyd)
Lloyd’s Register
Bureau Veritas
ABS (American Bureau of Shipping)
RINA (Registro Italiano Navale)
CCS (China Classi󹪝cation Society)
2 ... 25 Hz at 3.2 mm
25 ... 100 Hz at 40 m/s²
0.5 octave/min
1 frequency sweep/axis
Ampli󹪝cation factor (Q) < 2, 30 Hz/90 min
Ampli󹪝cation factor (Q) > 2, resonant frequency/90 min
1) Without bracket and aluminum enclosure
2) With mounting bracket, aluminum and stainless steel enclosure
3) Without mounting bracket, aluminum and stainless steel enclosure
8.4 Construction
Construction for gauge pressure and absolute pressure (from the gauge pressure series)
Weight Approx. 1.8 kg (3.9 lb) with aluminum enclosure
Approx. 3.8 kg (8.3 lb) with stainless steel enclosure
Material
Wetted parts materials
Technical data