Scale parameters and functions
8.3 DR 3 calibration parameters
SIWAREX WP321
Manual, 07/2014, A5E33715669A-02
51
8.3.12
Standstill range
Standstill monitoring checks whether the scales are correctly balanced. Scale standstill is
registered if the weight changes by less than a specified fluctuation in d (standstill value)
over a specified time (standstill time). Standstill monitoring is used in static scale mode
(commands: zeroing, taring). The diagram below illustrates how standstill monitoring works.
The current weight can only be registered in legal trade applications if standstill is reached.
Figure 8-1 Standstill monitoring
8.3.13
Standstill time
Standstill monitoring checks whether the scales are correctly balanced. Scale standstill is
registered if the weight changes by less than a specified fluctuation in d (standstill value)
over a specified time (standstill time). Standstill monitoring is used in static scale mode (with
the following commands: zeroing, taring).