Scale parameters and functions
8.3 DR 3 calibration parameters
SIWAREX WP321
Manual, 07/2014, A5E33715669A-02
51
8.3.12
Standstill range
Standstill monitoring checks whether the scales are correctly balanced. Scale standstill is
registered if the weight changes by less than a specified fluctuation in d (standstill value)
over a specified time (standstill time). Standstill monitoring is used in static scale mode
(commands: zeroing, taring). The diagram below illustrates how standstill monitoring works.
The current weight can only be registered in legal trade applications if standstill is reached.
Figure 8-1 Standstill monitoring
8.3.13
Standstill time
Standstill monitoring checks whether the scales are correctly balanced. Scale standstill is
registered if the weight changes by less than a specified fluctuation in d (standstill value)
over a specified time (standstill time). Standstill monitoring is used in static scale mode (with
the following commands: zeroing, taring).

Table of Contents

Other manuals for Siemens SIWAREX WP321

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens SIWAREX WP321 and is the answer not in the manual?

Siemens SIWAREX WP321 Specifications

General IconGeneral
BrandSiemens
ModelSIWAREX WP321
CategoryAccessories
LanguageEnglish

Related product manuals