Scale parameters and functions
8.9 DR 10 load cell parameters
SIWAREX WP321
68 Manual, 07/2014, A5E33715669A-02
8.9.1
50/60 Hz toggling
To improve the suppression of faults caused by the supply network, you can specify the
network frequency for signal filtering. The measuring rate is 100 Hz for the 50 Hz setting,
and 120 Hz for the 60 Hz setting.
8.9.2
Number of support points
If no anchor points are used, the number of support points is equal to the number of load
cells.
If anchor points are used in addition to load cells, the number of support points is equal to
the total number of load cells and fixed support points.
8.9.3
Load cell characteristic value
The load cell characteristic value is required to correctly interpret the output voltage from the
load cell. This specification is also necessary for determining load cell overload. The exact
value can be entered if the measurement log for the load cell is available. The mean value
can be entered if there is more than one load cell.
Example
Characteristic value = 2.018 mV/V
8.9.4
Rated load of a load cell
The rated load of a load cell is required for checking the maximum weighing range of the
scales. The rated load is entered in the specified units of weight.

Table of Contents

Other manuals for Siemens SIWAREX WP321

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens SIWAREX WP321 and is the answer not in the manual?

Siemens SIWAREX WP321 Specifications

General IconGeneral
BrandSiemens
ModelSIWAREX WP321
CategoryAccessories
LanguageEnglish

Related product manuals