2.2. INFRA D10 Micro Lead Acid battery
The short special battery cable included with the INFRA D10
Micro must be used for safety reasons (see chapter 8
Accessories).
Remove the battery from the INFRA D10 Micro before
mounting, demounting, carrying or transporting.
Use a nominal 12V, max 30Ah sealed Lead-Acid battery with the
INFRA D10 Micro
Use battery AGM or Gel type of batteries to minimize risk of
leakage.
Only use chargers provided or recommended by Sigicom, see
chapter 8 Accessories.
Under no circumstances, a charger which can supply more than
14.7V or 3A may be used
Never keep the battery in temperatures above 50 ºC (122 ºF)
If battery is shorted, it may burn the operator or generate sparks.
When charging the battery, in or out of the INFRA Datalogger,
place all parts in a well-ventilated room, away from open fire or
electric sparks.
If misused, the battery may generate inflammable gas, smoke or
leak sulfuric acid.
A worn-out battery, a battery with cracks or with signs to be
swollen or leaking should be replaced immediately.
The INFRA D10 Micro shall be mounted upright with the battery
standing at the bottom of the enclosure.
The battery strap included with the INFRA D10 Micro shall be
used to secure the battery.
If an external battery is used, the internal battery must be
disconnected.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sigicom INFRA D10 Micro and is the answer not in the manual?

Sigicom INFRA D10 Micro Specifications

General IconGeneral
BrandSigicom
ModelINFRA D10 Micro
CategoryData Loggers
LanguageEnglish

Related product manuals