Appendix C. Battery removal and installation
Battery removal
D10:
Remove the lower front panel by unscrewing the knurled screws.
Pull the batteries out of the battery holder.
If batteries are to be replaced, one battery at the time can be
replaced to keep the unit running while the replacement is done.
D10 Micro:
Use protective footwear when working with heavy batteries
Turn off the D10 Micro (see chapter 6.9.9 Power Off and
Reboot)
Release the strap holding the battery
Disconnect the short special battery cable from the longer battery
cable
Remove the battery from the enclosure
Battery installation
D10:
Remove the lower front panel by unscrewing the knurled screws.
Carefully, insert Sigicom B100 batteries into the battery holders
If batteries are to be replaced, one battery at the time can be
replaced to keep the unit running while the replacement is done.
Start the D10 and check the state of charge of the battery
D10 Micro:
Use protective footwear when working with heavy batteries

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sigicom INFRA D10 Micro and is the answer not in the manual?

Sigicom INFRA D10 Micro Specifications

General IconGeneral
BrandSigicom
ModelINFRA D10 Micro
CategoryData Loggers
LanguageEnglish

Related product manuals