38 SSA3000X User Manual
Amplitude
Edit the amplitude of the current point or line. If the Ref AMPT is enabled, edit the
amplitude difference between the amplitude of the current point and the reference
level.
Del Point
Delete the point you are editing.
Del All
Delete all point.
Save/Recall
Save or load the limit file.
2.2.6.3 Test
Enable or disable the limit test function.
2.2.6.4 Setup
Fail to stop
Select whether the instrument will continue or stop operation when a failure occurs.
Buzzer
Turn on or off the buzzer. When the buzzer is on, it beeps when a failure occurs.
X Axis
Set the X-axis unit to frequency or time unit.
Note that all the points of the current limit line will be deleted when the X-axis unit changes.
2.2.7 TG
Set the parameter related to the tracking generator (TG).
2.2.7.1 TG
TG is used to enable or disable the TG. When the TG is enabled, a signal with the same frequency
of the current sweep signal will be output from the [GEN OUTPUT 50Ω] connector at the front
panel. The power of the signal could be set through the menu.
2.2.7.2 TG Level
Set the output power of the signal of the tracking generator. You can use the numeric keys, knob or
direction keys to modify this parameter. For more details, please refer to “Parameter Setting”.
Table 2- 19 TG Level
Parameter
Explanation
Default
0 dB
Range
-200 dB ~ 200 dB
Unit
dB
Knob Step
1 dB
Direction Key Step
10 dB