Operating Manual SIGMA 2-16P page 23 of 75
04/2007
4 Set-up and Connection
4.1 Unpacking the Centrifuge
Open the cardboard box.
Take out the box containing the accessories.
Remove the upper foam cushion.
Lift the centrifuge upwards with a lifting device or with a sufficient number of
people to lift it safely. When lifting or carrying the centrifuge, always reach
under the centrifuge from the side.
) Attention!
The centrifuge weighs approx. 33 kg!
Please retain the packaging for any possible future transport of the centrifuge.
4.1.1 Transport Safety Device
The transport safety device of the SIGMA 2-16K centrifuge consists of a foamed
plastic ring in the centrifuge chamber. It must be removed prior to start-up.
Procedure:
Open the lid. To do so,
use the emergency release of the lid (see
6.2.5, page 45) if the centrifuge
is not connected to the power supply
or
connect the centrifuge to the power supply and press the lid key.
Unscrew the rotor tie-down screw.
Remove the foamed plastic ring from the rotor chamber, by lifting it carefully
on one side.
Please retain the transport safety device for the possibility of the return of the
centrifuge.