Operating Manual SIGMA 2-16P page 46 of 75
04/2007
6.2.6 Error Codes
Error no. Kind of error Measure Note
1-9 System error
• allow to slow down
• power off/on
All these errors stop
the centrifuge or cause
it to slow down.
10-19 Speedometer error
• allow to slow down
• power off/on
20-29 Motor error
• power off/on
• ensure ventilation
30-39 EEPROM error
• allow to slow down
• power off/on
With errors 34,35,36,
the centrifuge will stop.
With errors 37,38,
error message only.
40-45
Temperature error
(only for refrigerated
centrifuges)
• allow to slow down
• power off/on
• allow to cool down
• provide better ventilation
46-49
Imbalance error
• allow to slow down
• power off
• eliminate the imbalance
50-59 Lid error
• press lid key
• close lid
• remove foreign matter
from the opening of the
lid lock device
With error 50 and 51,
the centrifuge will stop.
60-69 Process error
• allow to slow down
• power off/on
With error 60,
message “power
failure during run”.
With error 61,
message “stop after
power on”.
70-79 Communication error
• allow to slow down
• power off/on
80-99 Parameter error
• allow to slow down
• provide for better
ventilation
• power off/on
With error 83, error
message only.
) Attention!
If it is impossible to eliminate the errors, please call the service!
6.2.7 Service Contact
Please contact your supplier for support or in the event of malfunctions and the
supply of spare parts:
SIGMA Laborzentrifugen GmbH
An der Unteren Söse 50
37520 Osterode
Service +49 55 22 50 07-8425 service@sigma-zentrifugen.de
Zentrale +49 55 22 50 07-0 info@sigma-zentrifugen.de
Fi
.6.2: Error Codes