EasyManua.ls Logo

Sigma 2-16P - 6.2.6 Error Code Reference; 6.2.7 Service Contact Information

Sigma 2-16P
75 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating Manual SIGMA 2-16P page 46 of 75
04/2007
6.2.6 Error Codes
Error no. Kind of error Measure Note
1-9 System error
allow to slow down
power off/on
All these errors stop
the centrifuge or cause
it to slow down.
10-19 Speedometer error
allow to slow down
power off/on
20-29 Motor error
power off/on
ensure ventilation
30-39 EEPROM error
allow to slow down
power off/on
With errors 34,35,36,
the centrifuge will stop.
With errors 37,38,
error message only.
40-45
Temperature error
(only for refrigerated
centrifuges)
allow to slow down
power off/on
allow to cool down
provide better ventilation
46-49
Imbalance error
allow to slow down
power off
eliminate the imbalance
50-59 Lid error
press lid key
close lid
remove foreign matter
from the opening of the
lid lock device
With error 50 and 51,
the centrifuge will stop.
60-69 Process error
allow to slow down
power off/on
With error 60,
message “power
failure during run”.
With error 61,
message “stop after
power on”.
70-79 Communication error
allow to slow down
power off/on
80-99 Parameter error
allow to slow down
provide for better
ventilation
power off/on
With error 83, error
message only.
) Attention!
If it is impossible to eliminate the errors, please call the service!
6.2.7 Service Contact
Please contact your supplier for support or in the event of malfunctions and the
supply of spare parts:
SIGMA Laborzentrifugen GmbH
An der Unteren Söse 50
37520 Osterode
Service +49 55 22 50 07-8425 service@sigma-zentrifugen.de
Zentrale +49 55 22 50 07-0 info@sigma-zentrifugen.de
Fi
g
.6.2: Error Codes

Table of Contents