2.4 Safety Instructions for Centrifugation
Ensure that the centrifuge was set up properly (see section 4.2 “Installation”,
page 23).
Check the centrifuge, rotor, and accessories for external signs of damage
prior to start-up.
Do not use the centrifuge with rotors and accessories that have not been
approved by the manufacturer. In case of doubt, contact our service team
(See 6.2.7 "Service Contact", page 46).
Ensure that rotor and buckets are correctly fitted (see 5.2.3 "Installation of
Rotors“, page 29 and Fig. 2.3).
Please observe the instructions on the installation of accessories (see 5.2.4,
page 31).
The load of the rotor as defined by the manufacturer and the maximum speed
must not be exceeded (see the engraving on the rotor or bucket).
The rotor must be loaded symmetrically at equal weights.
Fig. 2.3: Fitting of the buckets
1
2
1 Bucket installed correctly
2 Inadmissible operating state