Version 06/2007, Rev. 1.15 of 16/09/2011 - sb page 45 of 75
Translation of the original operating manual
6.2.6 Error Codes
Error no. Kind of error Measure Note
1-9 System error
allow to slow down
power off/on
All these errors stop
the centrifuge or cause
it to slow down.
10-19 Speedometer error
allow to slow down
power off/on
20-29 Motor error
power off/on
ensure ventilation
30-39 EEPROM error
allow to slow down
power off/on
With errors 34,35,36,
the centrifuge will stop.
With errors 37,38,
error message only.
40-45 Temperature error
(only for refrigerated
centrifuges)
allow to slow down
power off/on
allow to cool down
provide better ventilation
46-49 Imbalance error
allow to slow down
power off
eliminate the imbalance
50-59 Lid error
press lid key
close lid
remove foreign matter
from the opening of the
lid lock device
With error 50 and 51,
the centrifuge will stop.
60-69 Process error
allow to slow down
power off/on
With error 60,
message “power
failure during run”.
With error 61,
message “stop after
power on”.
70-79 Communication error
allow to slow down
power off/on
80-99 Parameter error
allow to slow down
provide for better
ventilation
power off/on
With error 83, error
message only.
Attention!
If it is impossible to eliminate the errors, please call the service!