Version 06/2007, Rev. 1.15 of 16/09/2011 - sb page 43 of 75
Translation of the original operating manual
6 Malfunctions and Error Correction
6.1 Error Mode
Malfunctions are indicated by error messages with a number in the speed and rcf
display. If the sound signal is activated, it sounds when the error message is
displayed (see 5.3.8 "Activating/Deactivating the Sound Signal", page 42).
6.2 Error Correction
Eliminate the source of the problem (see 6.2.1 to 6.2.6, page 43 ff).
Acknowledge the error messages with the lid key.
6.2.1 No Indication on the Display
Voltage in the socket? Check fuse in the mains supply.
Power cord plugged in and line
voltage present?
Plug in power cord correctly.
Fuse OK?
Switch unit off and let fuses cool down
for approx. 2 min.
Mains power switch on? Switch on power.
Lid closed? Close lid (see 6.2.4 “Lid cannot be
opened", page 44).
6.2.2 Centrifuge cannot be started
Start key LED not illuminated: Power off/on. If the error occurs again,
call the service.
Lid key LED flashes: Open and close the lid again. If error
occurs again although both locks are
locked, call the service.
6.2.3 Centrifuge decelerates during Operation
There was a brief mains power
failure (at least 2 sec), error
message 61
Centrifuge displays an error from
1 to 11 after powering on.
Press the start key in order to restart
the centrifuge.
Power off/on (see 6.2.6 "Error Codes",
page 45). If the error occurs again, call
the service.