Translation of the original operating manual
6.2.4 Lid cannot be opened
When trying to open the lid, the
lid lock has not released.
Close the lid again. Press down the
center of the lid until the lock audibly
locks. Open the lid again. If the error
occurs again, unlock the lid manually
(see 6.2.5 "Emergency lid release",
page 44) and call the service
The lid cannot be opened
although the lock has audibly
unlocked.
Check/clean the lid seal. Apply talcum
powder to the seal to avoid sticking.
6.2.5 Emergency Lid Release
In the event of a power supply failure, it is possible to manually open the
centrifuge lid using a stopper. It is located at the upper right side.
Switch off the mains power switch and disconnect the power cord from the
socket.
Remove the stopper, e.g. with a screwdriver.
Pull the visible string in order to release the lid.
Then, reinsert the stopper into the opening.
The lid may only be unlocked and opened when the rotor is at a standstill.
If the lid is opened by way of the emergency lid release during operation, the
centrifuge will be switched off immediately and decelerate brakeless.