4 Set-up and Connection
4.1 Unpacking the Centrifuge
The centrifuge is packaged in a slip-lid box.
• Remove the upper part of the box.
• Take out the box containing the accessories.
• Remove the upper foam cushion.
• Remove the slip-lid box.
• Lift the centrifuge upwards with a lifting device or with a sufficient number of
people to lift it safely. When lifting or carrying the centrifuge, always reach
under the centrifuge from the side.
Attention!
The centrifuge weighs approx. 48 kg!
Please retain the packaging for any possible future transport of the centrifuge.
4.1.1 Transport Safety Device
The transport safety device of this centrifuge consists of a foamed plastic ring in
the centrifuge chamber. It must be removed prior to start-up.
Procedure:
• Open the lid by pressing the lid key. If the centrifuge is not connected to the
power supply, use the emergency release of the lid (see 6.2.5).
• Unscrew the rotor tie-down screw.
• Remove the foamed plastic ring from the rotor chamber, by lifting it carefully
on one side.
• Please retain the transport safety device for the possibility of the return of the
centrifuge.