Sirona Dental Systems GmbH 6 Disinfection system
Care and cleaning instructions for the practice team C2
+
, M1
+
6.3 Sanitation
59 58 512 D 3370
D 3370.103.01.13.02
23
båÖäáëÜ
• Attach the suction hose (10) and the saliva ejector
hose and, if available, the 3rd suction hose (11) to
the side of the cleaning adapter (3).
When the sanitation bowl is completely fitted, briefly
press intensity key 1 to initiate the sanitation process.
(assistant element: briefly press the SAN key).
Sanitation phase 2:
First the program checks whether there is water flow at
the instrument couplings.
During this time, the SAN key on the assistant
element flashes approx. once a second.
Any instruments that are not withdrawn or do not con-
duct any water are excluded from the sanitation pro-
cess.
Error messages in sanitation phase 2
If there is no water flow, an acoustic signal sounds con-
tinuously and a message appears on the display to indi-
cate which instrument has no water flow. The instrument
number refers to the holders in ascending order from left
to right.
AE refers to the instrument holders on the dentist ele-
ment and HE to those on the assistant element.
The SAN key on the assistant element flashes approx. 4
times per second.
In this case open the water regulator of the correspond-
ing instrument on the instrument coupling. Leave the
corresponding instrument plugged into the adapter. As
soon as the system detects that water is flowing, the
sanitation process automatically continues.
However, if the acoustic signal sounds and you briefly
press intensity key 1 briefly without opening the water
regulator, the sanitation program continues without san-
itizing the corresponding water path (assistant element:
briefly press the SAN key).
Sanitation phase 3:
Then the water is pumped out of the water tank of the
water unit via the tumbler filler.