6 Disinfection system Sirona Dental Systems GmbH
6.3 Sanitation Care and cleaning instructions for the practice team C2
+
, M1
+
59 58 512 D 3370
24 D 3370.103.01.13.02
Sanitation phase 4:
The water tank is automatically filled with DENTOSEPT
P subsequently.
The SAN key on the assistant element lights up.
This activates flushing of the instrument hoses, the
SPRAYVIT hoses and the tumbler filler.
Sanitation phase 5:
The remaining sanitation time now appears on the
display, e.g. 23 hours and 57 minutes.
• Now the unit must be switched off for at least 24
hours, but not longer than three days (max. san-
itation time). Leave all instruments plugged into
their adapters.
If you turned off the treatment center water and air sup-
plies following sanitation phase 5, you must turn them on
again before using the treatment center again – i.e. after
24 hours have expired. Switch the unit back on after this
period. However, leave the instruments in the adapters of
the sanitation bowl.
• Switch the treatment center back ON after the mini-
mum sanitation time has expired and before the end
of the maximum sanitation time has been reached.
If an instrument is erroneously returned to its holder, the
following message appears:
PLACE ALL INSTR. IN SANIT. BOWL. The sanitation
program then automatically runs to its completion.
If you switch on the treatment center before the 24 hours
have elapsed (the remaining time is then indicated on
the display), you must briefly activate intensity key 1 for
> 3s after switching the unit on (assistant element: press
SAN key for 3 seconds). Pressing the key means abort-
ing sanitation (refer to section 6.3). The sanitation pro-
gram then runs to its completion.
Sanitation phase 6:
The instrument hoses, SPRAYVIT hoses and the tum-
bler filler are flushed with water. This takes several min-
utes.
During this time, the SAN key on the assistant
element flashes.