Sirona Dental Systems GmbH 6 Disinfection system
Care and cleaning instructions for the practice team C2
+
, M1
+
6.3 Sanitation
59 58 512 D 3370
D 3370.103.01.13.02
25
båÖäáëÜ
As soon as the message
PROCESS COMPLETE (
RETURN INSTR. (
appears on the display the instruments can be returned
to their holders.
The unit is now ready for operation again.
NOTE
i
If there is no assistant element on the water unit, no rins-
ing cycle occurs and it takes approx. 90 s until the mes-
sage
PROCESS COMPLETE ( appears.
Aborting the sanitation process
The sanitation process can be aborted in all phases by
pressing intensity key 1 for > 3s
(assistant element:press SAN key for 3 seconds).
Depending on the sanitation phase involved, sanitation
stops immediately, or the hoses are flushed with water
as described. Once the flushing process is complete and
the instruments have been returned to their holders, the
unit is ready for operation again.