Use and Operation
6-8
1 Reset
Reset the filament value, kV
feedback control time and
wireless module. The system
will shut down after reset.
N/A
2 Exposure Mode
Set Two Steps Exposure or One
Step Exposure
Two Steps Exposure
3 Filament Value
Filament value adjustment
range [0,255], the adjustment
step is 1.
Default filament
value
4 Exposure Count Count the number of exposures Not cleared
5
Timing of
Power Off
The default is 30 minutes.
The setting interval is [0,120].
Each adjustment is incremented
by 1 minute. When the
adjustment is 0, the power will
be turned off at an indefinite
time.
30 minutes
6
Timing of
Screen Blank
The default value is 600
seconds.
The setting interval is [30,1800],
and each adjustment is
increased by 30 seconds, and
the screen is not blank when the
adjustment is 0.
600 seconds
7
Timing of
Collimator
The default is 120 seconds.
The setting interval is [30, 300].
Each adjustment is incremented
by 10 seconds. It cannot be set
to 0.
120 seconds