15 PRESSURE WAVE
switch
Permits change of leading edge of pressure wave
from square to taper in 1-125 BPM range only.
16 ALARM MUTE
Pushbutton and
LED
Mutes audible alarms for one minute.
17 AIR OXYGEN
BLENDER control
Sets air/oxygen mix between 21 and 100% 0
2
±
3% with digital indication from independent
monitoring circuits.
18 O
2
inlet O
2
hose connector.
19 Medical air inlet Medical air hose connector.
20 Inspiratory
Pressure
Regulator
Adjusts driving pressure to set circuit inspiratory
pressure.
Range 0-60 cmH
2
O.
21 SYSTEM DRIVING
PRESSURE gauge
Approximate indication of the pressure above
PEEP which will be delivered to the patient in CMV,
PTV, SIMV or manual breath modes.
22 CPAP/PEEP
regulator
Sets CPAP level in the circuit, range is 0 to 15
cmH
2
O (nominal).
23 ALARM RESET
Pushbutton
Resets audible and visual alarms.
24 Fresh Gas Port 5LPM blended gas supply to patient (ventilator
powered up).
1LPM blended gas supply to patient (ventilator
OFF)
25 Proximal airway Proximal airway tube connector.
26 Exhalation block Connection for expiratory limb of patient circuit.
27 PIP, CYCLE FAIL
and HIGH alarm
LED’s with control
Sets visual and audible alarm level for inspiratory
pressure between 0 and 60 cmH
2
O.
28 CPAP Alarm LED
and control
Sets visual and audible alarm level for CPAP
pressure.