Do you have a question about the SLE 5000 and is the answer not in the manual?
Specifies the intended use of the ventilator for patients up to 20kg.
Continuous Positive Airway Pressure mode details and user settings.
Continuous Mandatory Ventilation mode details and user settings.
Patient Triggered Ventilation mode details and user settings.
Critical warnings that must be read and understood before using the ventilator.
General warnings applicable to all aspects of ventilator operation.
Specific operational warnings related to the ventilator unit itself.
General warnings and procedures for handling the flow sensor.
Procedure for testing the high and low oxygen alarms.
Instructions for calibrating the flow sensor.
Instructions for calibrating the oxygen sensor.
Setting Apnoea alarm delay and activating backup breaths in CPAP.
Setting Apnoea alarm delay for PTV, PSV, and SIMV.
Activating breath detection threshold in the flow waveform window.
Lists user-adjustable alarms and breath detection thresholds.
Initial preparation steps including cleaning and hose connection.
Connecting the patient circuit and bacterial filters to the ventilator.
Tests for mains power alarms without gas connected.
Tests related to oxygen and air gas supplies.
Procedure to test the high pressure alarm.
Final checks including oxygen calibration and readiness for use.
Explains the operation of high and low alarm thresholds for pressure waveforms.
Details how to set the breath detection threshold for flow triggering.
Describes breath trigger sensitivity when using pressure triggering without a flow sensor.
Instructions for setting the pressure trigger level in various modes.
Step-by-step instructions for calibrating the flow sensor.
Procedure for cleaning and sterilizing the N5402-REV2 flow sensor.
Explains the continuing positive pressure alarm and actions.
Lists ventilation problems, causes, and remedies.
Lists ventilator problems, causes, and remedies.
Procedure for performing a one-point oxygen calibration.
Procedure for performing a two-point oxygen calibration.
Requirement to perform functional tests periodically.
Summarizes the priority ratings and generation of alarms.
Details the Monitor Failure alarm and its actions.
Details the Sustained Sub-ambient Pressure alarm and its actions.
Details the Sub-Ambient Pressure alarm and its actions.
Details the Controller Failure alarm and its actions.
Details the Continuing Positive Pressure alarm and its actions.
Details the High Pressure alarm and its actions.
Details the Low Pressure alarm and its actions.
Details the Cycle Fail alarm and its actions.
Details the Pressure Sensor Drift alarm and its actions.
Details the No Gas alarm and its actions.
Details the No O2 Supply alarm and its actions.
Details the No Air Supply alarm and its actions.
Details the Battery Fault alarm and its actions.
Details the Battery Low alarm and its actions.
Details the Blocked Fresh Gas alarm and its actions.
Details the Leaking Fresh Gas alarm and its actions.
Details the Clean Flow Sensor alarm and its actions.
Details the Connect Flow Sensor alarm and its actions.
Details the High Minute Volume alarm and its actions.
Details the High Patient Leak alarm and its actions.
Details the Low Tidal Volume alarm and its actions.
Details the High Tidal Volume alarm and its actions.
Details the Low Minute Volume alarm and its actions.
Details the Breath Not Detected alarm and its actions.
Details the Apnoea (Volume) alarm and its actions.
Details the Apnoea (Pressure) alarm and its actions.
Details the Main Power Fail alarm and its actions.
Details the Fresh Gas Solenoid Fail alarm and its actions.
Details the Oxygen Cell Disconnected alarm and its actions.
Details the Calibrate Oxygen Cell alarm and its actions.
Details the Oxygen Cell Exhausted alarm and its actions.
Details the Oxygen Calibration Fail alarm and its actions.
Details the High Oxygen Level alarm and its actions.
Details the Low Oxygen Level alarm and its actions.
Details Monitor/Display Comms Fail alarm and its actions.
Details the Alarm Controller Fail alarm and its actions.
Details the Complete Power Fail Alarm and its actions.
Procedures for cleaning and disinfecting an in-service ventilator.
Lists and specifies ranges for user-settable alarms.
Tests the complete power fail alarm functionality.
Tests the backup alarm speaker by checking for beeps on parameter presses.
Steps to set O2% to 21% for testing and verify correspondence.
Procedure for functional testing when the flow sensor is not fitted.
Steps to calibrate the flow sensor, including actions for calibration failure.
Tests the high and low oxygen alarms by adjusting O2% levels.
Tests various functions and alarms in CPAP preview mode.
Steps to test the high pressure alarm by adjusting settings.
Steps to test the low pressure alarm by adjusting settings.
Steps to test the Breath Not Detected alarm.
Steps to test the Leak and Block alarms.
Tests the mains failure alarm by disconnecting power.
Tests gas supply alarms by disconnecting air and oxygen.
Tests the flow sensor disconnect alarm.
Tests the cycle fail alarm by adjusting its threshold.
Sets parameters and verifies oscillations in HFO Mode.
Tests the Pressure Change Detected alarm by disconnecting and reconnecting the flow sensor.
Specifies the intended use of the ventilator for patients up to 20kg.
Continuous Positive Airway Pressure mode details and user settings.
Continuous Mandatory Ventilation mode details and user settings.
Patient Triggered Ventilation mode details and user settings.
Critical warnings that must be read and understood before using the ventilator.
General warnings applicable to all aspects of ventilator operation.
Specific operational warnings related to the ventilator unit itself.
General warnings and procedures for handling the flow sensor.
Procedure for testing the high and low oxygen alarms.
Instructions for calibrating the flow sensor.
Instructions for calibrating the oxygen sensor.
Setting Apnoea alarm delay and activating backup breaths in CPAP.
Setting Apnoea alarm delay for PTV, PSV, and SIMV.
Activating breath detection threshold in the flow waveform window.
Lists user-adjustable alarms and breath detection thresholds.
Initial preparation steps including cleaning and hose connection.
Connecting the patient circuit and bacterial filters to the ventilator.
Tests for mains power alarms without gas connected.
Tests related to oxygen and air gas supplies.
Procedure to test the high pressure alarm.
Final checks including oxygen calibration and readiness for use.
Explains the operation of high and low alarm thresholds for pressure waveforms.
Details how to set the breath detection threshold for flow triggering.
Describes breath trigger sensitivity when using pressure triggering without a flow sensor.
Instructions for setting the pressure trigger level in various modes.
Step-by-step instructions for calibrating the flow sensor.
Procedure for cleaning and sterilizing the N5402-REV2 flow sensor.
Explains the continuing positive pressure alarm and actions.
Lists ventilation problems, causes, and remedies.
Lists ventilator problems, causes, and remedies.
Procedure for performing a one-point oxygen calibration.
Procedure for performing a two-point oxygen calibration.
Requirement to perform functional tests periodically.
Summarizes the priority ratings and generation of alarms.
Details the Monitor Failure alarm and its actions.
Details the Sustained Sub-ambient Pressure alarm and its actions.
Details the Sub-Ambient Pressure alarm and its actions.
Details the Controller Failure alarm and its actions.
Details the Continuing Positive Pressure alarm and its actions.
Details the High Pressure alarm and its actions.
Details the Low Pressure alarm and its actions.
Details the Cycle Fail alarm and its actions.
Details the Pressure Sensor Drift alarm and its actions.
Details the No Gas alarm and its actions.
Details the No O2 Supply alarm and its actions.
Details the No Air Supply alarm and its actions.
Details the Battery Fault alarm and its actions.
Details the Battery Low alarm and its actions.
Details the Blocked Fresh Gas alarm and its actions.
Details the Leaking Fresh Gas alarm and its actions.
Details the Clean Flow Sensor alarm and its actions.
Details the Connect Flow Sensor alarm and its actions.
Details the High Minute Volume alarm and its actions.
Details the High Patient Leak alarm and its actions.
Details the Low Tidal Volume alarm and its actions.
Details the High Tidal Volume alarm and its actions.
Details the Low Minute Volume alarm and its actions.
Details the Breath Not Detected alarm and its actions.
Details the Apnoea (Volume) alarm and its actions.
Details the Apnoea (Pressure) alarm and its actions.
Details the Main Power Fail alarm and its actions.
Details the Fresh Gas Solenoid Fail alarm and its actions.
Details the Oxygen Cell Disconnected alarm and its actions.
Details the Calibrate Oxygen Cell alarm and its actions.
Details the Oxygen Cell Exhausted alarm and its actions.
Details the Oxygen Calibration Fail alarm and its actions.
Details the High Oxygen Level alarm and its actions.
Details the Low Oxygen Level alarm and its actions.
Details Monitor/Display Comms Fail alarm and its actions.
Details the Alarm Controller Fail alarm and its actions.
Details the Complete Power Fail Alarm and its actions.
Procedures for cleaning and disinfecting an in-service ventilator.
Lists and specifies ranges for user-settable alarms.
Tests the complete power fail alarm functionality.
Tests the backup alarm speaker by checking for beeps on parameter presses.
Steps to set O2% to 21% for testing and verify correspondence.
Procedure for functional testing when the flow sensor is not fitted.
Steps to calibrate the flow sensor, including actions for calibration failure.
Tests the high and low oxygen alarms by adjusting O2% levels.
Tests various functions and alarms in CPAP preview mode.
Steps to test the high pressure alarm by adjusting settings.
Steps to test the low pressure alarm by adjusting settings.
Steps to test the Breath Not Detected alarm.
Steps to test the Leak and Block alarms.
Tests the mains failure alarm by disconnecting power.
Tests gas supply alarms by disconnecting air and oxygen.
Tests the flow sensor disconnect alarm.
Tests the cycle fail alarm by adjusting its threshold.
Sets parameters and verifies oscillations in HFO Mode.
Tests the Pressure Change Detected alarm by disconnecting and reconnecting the flow sensor.
| Type | High-Frequency Oscillatory Ventilator |
|---|---|
| FiO2 Range | 21 - 100% |
| Target Patient | Neonates, Infants, and Children |
| Ventilation Modes | HFO, CMV, SIMV |
| Respiratory Rate | 1 - 150 bpm |
| HFO Frequency | 5-15 Hz |
| Data Storage | Yes |
| Compatibility | Compatible with standard neonatal/pediatric circuits and accessories |
| Display | Touchscreen LCD |
| Channels | Pressure, Flow, Volume, Oxygen |