Page 157
Fresh gas solenoid fail
Oxygen too high
Oxygen too low
Oxygen cell not connected
Oxygen cell needs calibrating
Oxygen cell needs replacing
Oxygen calibration failure
Monitor/Display comms fail
Alarm controller failure
Complete power fail
30.5 Patient circuits
Volume of 10mm patient circuit without humidifying
chamber: 400ml
Resistance of 10mm circuit with 5 l/min fresh gas
flow = 0.35mbar (as measured from patient manifold
to atmosphere).
Resistance of 10mm circuit with 30 l/min fresh gas
flow = 10.0mbar (as measured from patient manifold
to atmosphere).
30.6 Outputs
Analogue: None
Digital Input/Outputs: Serial Port, IBM-AT-
compatible serial port, 9-pin
DSub, RS-232C
Available parameters:
Set BPM,
Set CPAP level
Set Tidal Volume
Set Insp Time
Set PIP
Set O
2
%
Set HFO Delta P
Set HFO Mean
Set HFO Rate
Ventilation Mode
TTV Status
Set Termination Sensitivity
Set Breath Trigger level
Set Breath Trigger threshold
Set Waveshape value
Set Patient Leak alarm value
Set Apnoea alarm value
Set Low Pressure alarm
threshold
Set Cycle Fail alarm
threshold
Set High Pressure alarm
threshold
Set Low Tidal Volume alarm
threshold
Set Low Minute Volume
alarm threshold
Set High Minute Volume
alarm threshold
Measured Total BPM
Measured CPAP
Measured Insp Time
Measured Insp Volume
Measured Exp Volume
Measured PEEP
Measured PIP
Measured O
2
%
Measured HFO Delta P
Measured HFO Mean
Trigger Count
Measured Minute Volume
Measured Leak
Measure Resistance
Measured Compliance
Measured C20/C
DCO
2
Current Alarm ID
For more details refer to the RS232 addendum.
Contact SLE for more details.
Compatibility with: Vuelink. Contact SLE for more
details.