25.3.3 Parameter Descriptions and Format
The text string outputted contains 41 parameters.
e.g.
60,2,6,10,23,100,4,2,100,1,20,0,45,20,30,160,280,
0,45,0,15000,60,3,10,145,139,3,25,99,22,13,0,824,
10,3275,6,39,280,64
A description of each parameter follows;
25.3.4 List of Parameters
Param.
Description Units Details
1 Set BPM breaths/minute 1 to 150
2 Set CPAP mbar Set CPAP pressure 0 to 20 mbar
3 Set Tidal Volume 0.2 ml Set Target Tidal volume 10 to 1000 (2 to
200ml)
4 Set Insp. Time 0.01 sec. Set insp. time or max insp time, 10 to 300
(0.10 to 3.00 seconds)
5 Set PIP mbar Set PIP pressure 0 to 65 mbar
6Set O
2
% % Set O2 Concentration 21 to 100%
7 Set HFO Delta P mbar Set HFO Delta P 4 to 180 mbar
8 Set HFO Mean mbar Set mean 0 to 45 mbar
9 Set HFO Rate 0.1 Hz Set HFO rate 30 to 200 (3 to 20Hz)
10 Ventilation Mode None 30 = Ventilation OFF
0 = CPAP
1 = CMV
31 = PTV
32 = PSV
3 = SIMV
4 = HFO Only
5 = HFO + CMV
11
TTV
plus
Status
None 0 = off
255 = on
12 Set Termination Sensitivity % Set percentage
0 to 50 = percentage of max. flow that
triggers termination,
if value < 0 then termination sens. is OFF
13 Set Breath Trig. Threshold l/min for flow
triggering.
No units for
pressure
triggering.
2 to 200 (0.2 to 20 l/min for flow triggering).
14 Set Waveshape None Set Rise time value 0 to 25 (rise time is
proportional to the value)