30.1.8 Controls (Via touchscreen display)
Adjust button: Activates alarm thresholds for
modification
Alarm Auto-Track
button:
Sets selected % O
2
for
testing
Apnea Sup Button: Sends diagnostic pulse
Apnoea setup
button:
Activates Backup/Apnea
settings panel
Arrow buttons: Increments value
Auto set Button: Autoset high, cycle and low
alarm thresholds
Backup parameter
control:
1 to 150 bpm (Resolution 1
bpm)
Begin psv Button: Sends diagnostic pulse
BPM parameter
control:
1 to 150 bpm (Resolution 1
bpm)
Calibrate button: Calibrates flow sensor
Calibrate Flow
Sensor button:
Activates calibrate flow
sensor panel
Camera button: Stores a loop to memory
CMV button: Selects CMV mode
Confirm button: Confirms selection
Continue without
flow button:
Selects pressure monitoring
operation
Continuous HFO
button:
Selects oscillations in both
inspiratory and expiratory
phases.
Controller button: Activates controller services
panel
CPAP button: Selects CPAP mode
CPAP parameter
control:
0 to 20 mbar (Resolution 1
mbar)
Delta P parameter
control:
4 to 180 mbar (Resolution 2
mbar)
Default button Returns the waveform display
to default layout.
End psv button: Sends diagnostic pulse
Expiratory Only
button:
Selects oscillations in
expiratory phase only
F/P button: Selects flow versus pressure
loop display
F/V button: Selects flow versus volume
loop display
O
2
button: Selects fractional
concentration of inspired
oxygen against time display
O
2
parameter
control:
21% to 100% (Resolution 1
%)
Flow button: Activates flow sensor
calibration panel
Folder button: Retrieves loop from memory
Graphs button: Activates waveform and
loops panels
HFO activity
button:
Activates HFO activity panel
HFO I:E Selects I:E ratio 1:1/1:2/1:3
HFO Only button: Selects HFO only mode
HFO rate
parameter control:
3 to 20 Hz (Resolution 1 Hz)
HFO+CMV button: Selects HFO+CMV mode
Hist button: Display alarm history panel
Language button: Change interface language
Manual Breath
button:
Delivers one manual breath
Mean parameter
control:
0 to 45 mbar (Resolution 1
mbar)
Min Vol button: Selects minute volume
against time display
Mode select
button:
Activates mode select panel
Monitor button: Activates monitor services
panel
Monitor Pressure
Button:
Toggles between waveforms,
diagnostic use only
Mute button: Mutes a mutable alarm for 1
minute.
Next button: Activates next panel
Night and screen
lock button:
Activates night mode for LCD
ON/OFF button: Toggles function state
between on and off
Options and
service data
button:
Activates calibration options
panel
Oxygen alarm test
button:
Activates Oxygen Alarm
testing panel
Oxygen button: Activates oxygen sensor
calibration panel
P Max button: Selects maximum pressure
against time display
P Mean button: Selects mean pressure
against time display
P Min button: Selects minimum pressure
against time display
Pause/Play button: Freezes & unfreezes
waveform display (60 secs
max)