Page 152
30. Technical Specification
This section summarizes the specification of the
ventilator in terms of the modes, ranges and limits
that are required on the controls and the displays. It
also summarizes the mechanical and electrical
power constraints.
30.1 Operating Modes Conventional
Ventilation
30.1.1 CPAP
Inspiratory Time: 0.1 to 3.0 seconds
(Resolution 0.01 seconds)
CPAP Pressure: 0 mbar to 20 mbar
(Resolution 1mbar)
Inspiratory Pressure: 0 mbar to 65 mbar
(Resolution 1mbar)
Volume Targeting: 2 ml to 200 ml
From 2ml to 10ml the parameter
increments in 0.2ml steps (Fine
resolution).
From 10ml to 100ml the parameter
increments in 1ml steps (Standard
resolution).
From 100ml to 200ml the parameter
increments in 5ml steps (Coarse
resolution).
O
2
%: 21% to100% (Resolution 1%)
Backup breaths are pressure limited, time cycled.
30.1.2 CMV
BPM: 1 to 150 (Resolution 1BPM)
I:E Ratio: Calculated from BPM and
Inspiratory time settings.
(11.2:1 to 1:600)
Inspiratory Time: 0.1 to 3.0 seconds
(Resolution 0.01 seconds)
PEEP Pressure: 0 mbar to 20 mbar
(Resolution 1mbar)
Inspiratory Pressure: 0 mbar to 65 mbar
(Resolution 1mbar)
Volume Targeting: 2 ml to 200 ml
From 2ml to 10ml the parameter
increments in 0.2ml steps (Fine
resolution).
From 10ml to 100ml the parameter
increments in 1ml steps (Standard
resolution).
From 100ml to 200ml the parameter
increments in 5ml steps (Coarse
resolution).
O
2
%: 21% to 100% (Resolution
1%)
Breaths can be pressure limited, time cycled.
30.1.3 PTV
Inspiratory Time: 0.1 to 3.0 seconds
(Resolution 0.01 seconds)
CPAP Pressure: 0 mbar to 20 mbar
(Resolution 1mbar)
Inspiratory Pressure: 0 mbar to 65 mbar
(Resolution 1mbar)
Volume Targeting: 2 ml to 200 ml
From 2ml to 10ml the parameter
increments in 0.2ml steps (Fine
resolution).
From 10ml to 100ml the parameter
increments in 1ml steps (Standard
resolution).
From 100ml to 200ml the parameter
increments in 5ml steps (Coarse
resolution).
O
2
%: 21% to 100% (Resolution
1%)
Backup breaths are pressure limited, time cycled.
30.1.4 PSV
Inspiratory Time: 0.1 to 3.0 seconds
(Resolution 0.01 seconds)
CPAP Pressure: 0 mbar to 20 mbar
(Resolution 1mbar)
Inspiratory Pressure: 0 mbar to 65 mbar
(Resolution 1mbar)
Volume Targeting: 2 ml to 200 ml
From 2ml to 10ml the parameter
increments in 0.2ml steps (Fine
resolution).
From 10ml to 100ml the parameter
increments in 1ml steps (Standard
resolution).
From 100ml to 200ml the parameter
increments in 5ml steps (Coarse
resolution).
O
2
%: 21% to 100% (Resolution
1%)
Backup breaths are pressure limited, time cycled.