30.1.9 Controls
30.2 Measurement
30.2.1 Flow and Volume
Flow Sensor Type: 10 mm dual-hot-wire-
anemometer with
replaceable hot-wire
subassembly, autoclaveable.
Sensor electrically isolated.
Flow Rate: 0.2 l/min to 32 l/min
(Accuracy ±8 % maximum)
Expiratory Tidal Volume:
0 to 999ml (0.1ml)
Expiratory Minute Volume:
0 to 18L (Resolution: 1ml)
Deadspace: 1 ml
Weight: 10 g
Conventional Ventilation and combined modes only:
Tube Leakage: 0 to 99% (Resolution: 1%)
Averaged over 10 breaths
Breath Rate (total): 0 to 150 BPM (Resolution:
1BPM)
Dynamic Compliance: 0 to 100ml/mbar (Resolution:
1ml/mbar)
C20/C: Resolution 0.1
Sampling Time: 2ms
Resistance: 0 to 1000 mbar.second/l
Triggering: Inspiratory flow (0.2 to 10 l/
min)
Above values are obtained under ATPD (ambient
temperature and pressure,
dry) conditions.
HFO only:
DCO
2
: Resolution 1
30.2.2 Oxygen Concentration
Range: 21% to 100% (Resolution
1%, accuracy ±2%)
PEEP parameter
control:
0 to 20 mbar (Resolution 1
mbar)
PIP parameter
control:
0 to 65 mbar (Resolution 1
mbar)
Picture capture
button
Stores loop to memory
PSV button: Selects PSV mode
PSV Setup: button: Activates pressure support
panel
PTV button: Selects PTV mode
Reset button: Resets current alarm
Reset
contamination
alarm button:
Resets the alarm
Return button: Returns to previous panel
Select button: Tabs between time and date
values
Set BPM button: Selects BPM against time
display
Set time and date
button:
Activates time and date panel
Set Trigger button Activates breath detection
trigger threshold
SIMV button: Selects SIMV mode
Standby button: Sets ventilator into 90 second
standby mode
Stored loop button Displays stored loops
Ti parameter:
control:
0.1 to 3 seconds (Resolution
0.01 seconds)
Tidal vol
parameter: control:
2 to 200 ml
From 2ml to 10ml the parameter
increments in 0.2ml steps (Fine
resolution).
From 10ml to 100ml the parameter
increments in 1ml steps (Standard
resolution).
From 100ml to 200ml the parameter
increments in 5ml steps (Coarse
resolution).
TTV button: Selects TTV on/off panel
Two point O
2
calibration button:
Activates calibrates O
2
system
Ventilation Off
button:
Selects ventilation off mode
Version button: Activates version information
panel
Vmin Selects minute volume
against time display
V/P button: Selects volume versus
pressure loop display
Vte button: Selects tidal volume against
time display
Waveform display
sync button:
Toggles waveform display
synchronization
Waveshaping
button:
Activates change wave shape
panel
Zero button: Zeros seconds
Zoom in button Zooms in on a trend display
Zoom out button Zooms out of a trend display
ON/OFF: Control for power supply