Page 123
21.2 Ventilator mounting.
The ventilator needs to be secured to the trolley with
the screws and washers provided in the fittings
pack.
Use two M4 x10mm screws and two M4 shake proof
washers to secure the ventilator to the trolley.
21.3 Back-up battery connection
The internal back-up batteries need to be connected
prior to use. Remove the rear cover to reveal the
battery fuse holder.
Open the fuse holder and move the fuse from the
front compartment to the rear.
Close the compartment and re fit the outer cover.
21.4 Ventilator labelling
Refer to section 35. on page 184 “Ventilator
labelling” on where to apply the labels.
Note: Clean the area with the alcohol wipe
provided, prior to application of the label.