Page 101
17.2 Cleaning and Sterilization of the
N5402-REV2 Sensor
Remove the flow sensor connecting cable before
any cleaning, disinfection or sterilization.
Rinse the sensor body immediately after use and
put it into a disinfectant (recommended by the
infection control authority of the hospital/
organization), otherwise the sensor will encrust and
cannot be used any more.
Prior to first use as well as after each use Clean or
Disinfect/Sterilize the sensor.
Cleaning: A soap solution or mild alkaline
solution should be used.
Disinfection: Use commercially available
disinfectants that are recommended for
use with PLASTIC MATERIALS.
Immersion times and concentrations
stated must be in accordance with
manufacturer’s instructions.
Sterilization For sterilization a validated steam
sterilization according to BS EN554
can be applied. Autoclave at
134ºC (277ºF) (Allowable variation of
temperature of +3ºC) at 220kPa
(32psi) with a minimum holding time of
3 minutes.
or
121ºC (248ºF) (Allowable variation of
temperature of +3ºC) at 96kPa
(14.1psi) with a minimum holding time
of 15 minutes.
Before each use the sensor must be checked for
damage. Damaged parts must not be used.
Warning: Do not clean the flow sensor with
compressed air or water jet. As this will
destroy the sensor wires.
Note: Disinfectants containing compounds
similar to PHENOL or ALKYLAMINES
(Glucorrotamine) are unsuitable.
Note: Eliminate all residues of cleaning
agents and disinfectant used by thoroughly
rinsing with sterile water after each cleaning
and disinfecting procedure.
Flow sensor body
ET tube port
ET Manifold port
Flow sensor
connecting cable
(Do not autoclave
or immerse in
liquids)
Note: The sensor must not be connected to
other standard connectors when sterilised to
prevent cracking. Ensure no other
components/items are lying on the sensor
during the sterilization process.
Warning: Do not use any nebulized gases
(medications, salt solutions, etc.) in
conjunction with the sensor as they are likely
to degrade the performance of the sensor and
subsequent displayed accuracies.