EasyManua.ls Logo

SLE 5000 - Ventilator Specifications Continued; SIMV Mode Specifications; HFO Ventilation Specifications; HFO+CMV Mode Specifications

SLE 5000
200 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 153
30.1.5 SIMV
BPM: 1 to 150 (Resolution 1BPM)
I:E Ratio: Calculated from BPM and
Inspiratory time settings.
(11.2:1 to 1:600)
Inspiratory Time: 0.1 to 3.0 seconds
(Resolution 0.01 seconds)
PEEP Pressure: 0 mbar to 20 mbar
(Resolution 1mbar)
Inspiratory Pressure: 0 mbar to 65 mbar
(Resolution 1mbar)
Volume Targeting: 2 ml to 200 ml
From 2ml to 10ml the parameter
increments in 0.2ml steps (Fine
resolution).
From 10ml to 100ml the parameter
increments in 1ml steps (Standard
resolution).
From 100ml to 200ml the parameter
increments in 5ml steps (Coarse
resolution).
O
2
%: 21% to 100% (Resolution
1%)
Breaths are flow cycled, pressure limited. Backup
breaths are time cycled pressure limited.
30.1.6 HFO Ventilation
HFO Only
Frequency Range: 3-20Hz. (1Hz resolution.)
I:E Ratio 1:1 / 1:2 / 1:3
Delta Press Range: 4 mbar to 160 mbar for 3.3 to
6 bar inlet gas pressures
4 mbar to 140 mbar 2.8 to 3.2
bar inlet gas pressures
(Resolution 2mbar)
Mean airway range: 0 mbar to 45 mbar maximum
(Resolution 1mbar)
O
2
%: 21% to 100% (Resolution
1%)
30.1.7 HFO+CMV
BPM: 1 to 150 (Resolution 1BPM)
Inspiratory Time: 0.1 to 3.0 seconds
(Resolution 0.01 seconds)
Frequency Range: 3-20Hz. (1Hz resolution.)
I:E Ratio Calculated from BPM and
Inspiratory time settings.
(11.2:1 to 1:600)
Inspiratory Pressure: 2 mbar to 65 mbar
(Resolution 1mbar)
Delta Press Range: 4 mbar to 180 mbar 3.3 to 6
bar inlet gas pressures
4 mbar to 140 mbar 2.8 to 3.2
bar inlet gas pressures
(Resolution 2mbar)
PEEP: 0 mbar to 20 mbar maximum
(Resolution 1mbar)
O
2
%: 21% to 100% (Resolution
1%)

Table of Contents

Related product manuals