30.2.3 Pressure
Real Time Pressure: Resolution 1mbar, accuracy
±1mbar
Drift detection: detects drift when there is a
pressure difference of more
than 5mbar between the two
pressure transducers. Only
active when the measured
pressure is below 70mbar.
Sampling Time: 2ms
Peak Pressure: 0 to 175 mbar (resolution
1mbar, accuracy ±0.75% of
full scale range)
PEEP Pressure: 0 to 175 mbar (resolution
1mbar, accuracy ±0.75% of
full scale range)
Mean Pressure: -175 to 175 mbar (resolution
1mbar, accuracy ±0.75% of
full scale range)
Delta P In HFO combined mode
Delta P is measured during
expiration only.
30.3 Exhalation Block Port Jet Sizes
Reverse or negative jet: Ø 1.45/1.55 mm
Forward or positive jet: Ø 1.25/1.35 mm
Mean or 3rd jet: Ø 0.60/0.62 mm
30.4 Alarms
30.4.1 User settable Alarms
High Pressure:
Autoset when patient pressure controls are
adjusted, or manually
adjustable.
Range: 10 to 110 mbar
Resolution: 0.5 mbar
Cycle Fail:
Autoset when patient pressure controls are
adjusted, or manually
adjustable.
Range: 0 or 5 mbar above low-
pressure threshold
(whichever is lower) to 5
mbar below high-pressure
threshold.
Resolution: 0.5 mbar
Low Pressure:
Autoset when patient pressure controls are
adjusted, or manually
adjustable.
Range: -10 mbar (Conventional)/ -70
mbar (HFO modes) to 10
mbar below high-pressure
threshold.
Resolution: 0.5 mbar
High Tidal Volume:
Range: 3 to 200 ml
Resolution: 0.2 ml
Low Tidal Volume:
Range: 0 to 200 ml
Resolution: 0.2 ml
Low Minute Volume:
Range: 0 to 0.02 litres below High
minute volume threshold.
Resolution: 0.1 litre
High Minute Volume:
Range: 0.02 to 18 litres.
Resolution: 0.1 litre
Apnoea time:
Range: 5 to 60 seconds
Resolution: 1 second
30.4.2 Obligatory Alarms (non adjustable by
user)
Monitor Failure
Sustained sub ambient
Sub ambient
Controller Failure
System fail, Memory checksum error
System fail, Alarm controller error
Gas not connected
O2 not connected
Air not connected
Block alarm
Leak alarm
Battery fault
Battery low
Continuing positive pressure
Pressure sensor drift
Isolate system error
Flow ADC unable to calibrate
Serial communication error
Flow sensor defect
Clean flow sensor
Flow sensor not connected
Unable to calibrate flow sensor
Flow sensor not calibrated
User interface failure
Patient Leak
Pressure change detected
Mains failure