Page 97
16.8.2 Interactive and limiting controls in HFO.
The user needs to be aware that the following
controls interact.
The HFO I:E ratio & the Delta P control interact.
When changing the HFO I:E Ratio the set Delta P
can be reduced when the ratio is increased.
The Mean & the Delta P control interact. When
changing the mean the set Delta P can be reduced
when the mean is increased.
16.8.3 Ventilation Without a Flow Sensor
Connected.
When the ventilator is used without the flow sensor
connected flow based functions, waveforms and
loops will not be displayed.
Note: Decreasing the HFO I:E ratio will NOT
increase the Delta P.
Note: Decreasing the Mean setting will NOT
increase the Delta P.