SLE 5000 - Interactive and Limiting Controls in HFO; Ventilation Without a Flow Sensor Connected

SLE 5000
200 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 97
16.8.2 Interactive and limiting controls in HFO.
The user needs to be aware that the following
controls interact.
The HFO I:E ratio & the Delta P control interact.
When changing the HFO I:E Ratio the set Delta P
can be reduced when the ratio is increased.
The Mean & the Delta P control interact. When
changing the mean the set Delta P can be reduced
when the mean is increased.
16.8.3 Ventilation Without a Flow Sensor
Connected.
When the ventilator is used without the flow sensor
connected flow based functions, waveforms and
loops will not be displayed.
Note: Decreasing the HFO I:E ratio will NOT
increase the Delta P.
Note: Decreasing the Mean setting will NOT
increase the Delta P.