Page 145
27. Bacterial Filters
It is recommended that bacterial filters are fitted in
the fresh gas supply and on the patient side of the
exhalation block.
The filters reduce the possibility of infection to the
patient and contamination of the ventilator from
secretions or fluids in the breathing circuits that
could accidentally enter the ventilators gas ports.
The ventilator can be used without bacterial filters in
place, but the user must take extra care in not
allowing secretions or fluids to enter the ventilators
gas ports.
27.1 Bacterial filter, SLE Part Nº:N2029
(Autoclavable)
This autoclavable bacterial filter is fitted into the
humidifier supply line and has to be fitted in
accordance with the indicator arrow embossed on
the surface of the filter.
Autoclave with pure dry saturated steam at:
134ºC (277ºF) (Allowable variation of temperature
of +3ºC) at 220kPa (32psi) with a minimum holding
time of 3 minutes
or
121ºC (248ºF) (Allowable variation of temperature
of +3ºC) at 96kPa (14.1psi) with a minimum holding
time of 15 minutes.
The filter can be autoclaved a maximum of 25 times
within its anticipated service life of 12 months.
For other makes of bacterial filter please refer to
manufacturers instructions.
27.2 Bacterial filter, SLE Part Nº: N2587/
000/001 (Single use)
This single use bacterial filter is fitted onto the
exhalation block outlet. This filter should be
disposed of in accordance with local hospital
authority guidelines. A new filter should be used
with each patient.
For other makes of bacterial filter please refer to
manufacturer’s instructions.
Warning: Do not immerse the filter in any
liquid. If the filter is allowed to become wet it
will no longer function correctly and should
be discarded.
Bacterial Filter
Bacterial Filter
(SLE N
o
:N2587/000/001)
(SLE N
o
:N2029)
Cautions: When Using Bacterial Filter
N2587/000/001
The user should be aware that any occlusion
(ingress of H
2
O) of the filter increases the
resistance to airflow, resulting in increased or
erratic airway pressures. Airway pressures
should be monitored during use and the filter
changed if found to be contaminated in any
way. When using humidification, the filter
should be checked regularly for signs of
water build up which could cause occlusion.
The filter must be changed every 24 hours, or
when there is evidence of water in the filter.