Page 95
16.7.4 Ventilation Without a Flow Sensor
Connected.
When the ventilator is used without the flow sensor
connected, the user will have to set the Breath
Trigger Sensitivity and the Cycle Fail alarm.
Flow based functions, waveforms and loops will not
be displayed.
TTV
plus
will not be available.
Refer to chapter "15.2 Breath Trigger Sensitivity
(Pressure Triggering)" on page 80 for more
information.
Note: If TTV
plus
is ON the termination
sensitivity and provide support values for
non SIMV breaths will only come into effect if
the targeted tidal volume could not be
reached.