Page 178
32. EMC compliance
The SLE4000/5000 electro magnetic compatibility
was tested to and meets the requirements of the
following relevant standards:
EN60601-1-2
EN61000-3-2
EN61000-3-3
Guidance and manufacturer's declaration - electromagnetic emissions
The SLE5000 is intended for use in the electromagnetic environment specified below. The customer or the
user of the SLE5000 should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment- guidance
RF emissions
CISPR 11
Group 1 The SLE4000/5000 uses RF energy only for its internal func-
tion. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equip-
ment.
RF emissions
CISPR 11
Class B
The SLE4000/5000 is suitable for use in all establishments
other than domestic, and may be used in domestic
establishments and those directly connected to the public
low-voltage power supply network that supplies buildings
used for domestic purposes, provided the following warning
is heeded:
Warning: This equipment/system is intended for use by
healthcare professionals only. This equipment/ system may
cause radio interference or may disrupt the operation of
nearby equipment. It may be necessary to take mitigation
measures, such as re-orienting or relocating the SLE4000/
5000 or shielding the location.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies