Page 130
15 Set Patient Leak Alarm % Percentage leak at which alarm occurs.
10 to 50%.
Values above 50 = alarm off.
16 Set Apnoea Alarm seconds 5 to 60 seconds
17 Set Low Pressure Alarm 0.1 mbar -1200 to 1100 (-120 to 110 mbar)
18 Set Cycle Fail Alarm 0.1 mbar 0 to 1150 (0 to 115 mbar)
19 Set High Pressure Alarm 0.1 mbar 100 to 1200 (10 to 120 mbar)
20 Set Low Tidal Vol. Alarm 0.1 ml 0 to 2970 (0 to 297ml)
21 Set High Tidal Vol. Alarm 0.1 ml 30 to 3000 (3 to 300ml)
22 Set Low Minute Vol. Alarm ml 0 to 18000 (0 to 18l)
23 Set High Minute Vol. Alarm ml 1000 to 18000 (1 to 18l)
24 Measured Total BPM breaths/min Total breath count over the last minute 0 to
255.
25 Measured CPAP mbar -175 to +175 mbar
26 Measured Insp. Time 0.01 sec. 0 to 32768 (0 to 327.68 seconds)
27 Measured Insp. Volume 0.1ml 0 to 32768 (0 to 3.2768 l)
28 Measured Exp. Volume 0.1ml 0 to 32768 (0 to 3.2768 l)
29 Measured PEEP mbar -175 to 175 mbar
30 Measured PIP mbar -175 to 175 mbar
31 Measured O
2
% % 0 to 100 %
32 Measured HFO Delta P mbar -175 to 175 mbar
33 Measured HFO Mean mbar -175 to 175 mbar
34 Trigger Count breaths/min Number of patient triggers in the last
minute 0 to 255
35 Measured Minute Vol. ml 0 to 32768 (0 to 32.768l)
36 Measured Leak % 0 to 100% leak
37 Measured Resistance 0.1 mbar.sec/l Airway resistance 0 to 32768 (0 to 3276.8
millibar seconds per litre)
38 Measured Compliance 0.1 ml/mbar Total Compliance 0 to 255 (0 to 25.5
milliliters per millibar)
39 Measured C20/C 0.1 This value is a ratio of the compliance over
the last 20% of the pressure rise compared
to the total compliance. 0 to 255 (0 to 25.5)
40 DCO
2
1 0 to 32000
41 Current Alarm None This value indicates the currently displayed
alarm condition (see table opposite)
Param.
Nº
Description Units Details