Alarm 19. Unexpected Rise in Max P
Alarm message.................... Unexpected Rise in
Max P.
Alarm sub message............. Press AUTOSET to
adjust HFO alarms
to new pressures
Priority of alarm ................... 19
Monitor mode....................... HFO Only
Can alarm be muted ........... Yes
Sounder Priority................... High
Alarm Description:
When the ventilator set alarm threshold of 5mbar
above the max pressure is crossed, this alarm is
generated.
Action: Check patient. Press autoset or adjust
ventilation parameters.
Alarm 20. Unexpected Drop in Max P
Alarm message.................... Unexpected Drop in
Max P.
Alarm sub message............. Press AUTOSET to
adjust HFO alarms
to new pressures
Priority of alarm ................... 20
Monitor mode....................... HFO Only
Can alarm be muted ........... Yes
Sounder Priority................... High
Alarm Description:
When the ventilator set alarm threshold of 5mbar
below the max pressure is crossed, this alarm is
generated.
Action: Check patient. Press autoset or adjust
ventilation parameters. Check patient circuit
& water trap.
Alarm 21. Unexpected Rise in Min P
Alarm message.................... Unexpected Rise in
Min P.
Alarm sub message ............. Press AUTOSET to
adjust HFO alarms
to new pressures
Priority of alarm.................... 21
Monitor mode....................... HFO Only
Can alarm be muted ........... Yes
Sounder Priority ................... High
Alarm Description:
When the ventilator set alarm threshold of 5mbar
above the min pressure is crossed this alarm is
generated.
Action: Check patient. Press autoset or adjust
ventilation parameters.
Alarm 22. Unexpected Drop in Min P
Alarm message.................... Unexpected Drop in
Min P.
Alarm sub message ............. Press AUTOSET to
adjust HFO alarms
to new pressures
Priority of alarm.................... 22
Monitor mode....................... HFO Only
Can alarm be muted ........... Yes
Sounder Priority ................... High
Alarm Description:
When the ventilator set alarm threshold of 5mbar
below the min pressure is crossed, this alarm is
generated.
Action: Check patient. Press autoset or adjust
ventilation parameters.